Validated BP Monitors for For Specialist Use
Manufacturer | Model | Type | Price | Protocol | Validation Study | Features | Notes | Reference |
---|---|---|---|---|---|---|---|---|
A C Cossor & Son | Greenlight 300 | Clinic | Over £100 | International Protocol | Independent | 4 x AA batteries | Manual Device | Graves JW, Tibor M, Blaithnead M, Klein L & Sheps SG. The Accoson Greenlight 300, the first non-automated mercury-free blood pressure measurement device to pass the International Protocol for blood pressure measuring devices in adults. Blood Pressure Monitoring 2004, 9(1):13-17 |
A&D | TM-2430 | ABPM | Over £100 | BHS A/A | Independent | Software included. Data output facility to PC or printer. 3 x AA NiMH rechargeable (included) or AA Alkaline | Palatini P, Frigo G, Bertolo O et al. Validation of the A&D TM-2430 device for ambulatory blood pressure monitoring and evaluation of performance according to sujects’ characteristics. Blood Pressure Monitoring 1998: 3:255-260 | |
A&D | TM2440 (Derivative of TM2441) | ABPM | Over £100 | ISO 81060-2:2013 | Independent | ABPM device | Kazuomi Kario, Satoshi Hoshide, Kimiyo Saito, Keiko Sato, Haruna Hamasaki, Hiromi Suwa and Naoko Tomitani Validation of the TM-2441 ambulatory blood pressure measurement device according to the ISO 81060-2:2013 standard Blood Pressure Monitoring 2019, 24:38–41 |
|
A&D | TM2441 | ABPM | Over £100 | ISO 81060-2:2013 | Independent | ABPM device | Kazuomi Kario, Satoshi Hoshide, Kimiyo Saito, Keiko Sato, Haruna Hamasaki, Hiromi Suwa and Naoko Tomitani Validation of the TM-2441 ambulatory blood pressure measurement device according to the ISO 81060-2:2013 standard Blood Pressure Monitoring 2019, 24:38–41 |
|
A&D | TM-2655 | Waiting Room | Over £100 | BHS A/A | Independent | Mains Powered | Validated for self-assessment in pharmacies etc | Kobalava ZD, Ktoovskaya YV, Babaeva LA & Moiseev VS. Validation of TM-2655 oscillometric device for blood pressure measurement. Blood Pressure Monitoring 2006, 11 (2):87-90 |
A&D | TM-2656 | Waiting Room | Unknown Price | BHS A/A | Independent | Static, automatic arm cuffing system suitable for use in clinical settings for patients to self-measure BP. | Wei-Fang Z, Ming L, Yuan-Yuan K, Yan L and Ji-Guang W. Validation of the fully automated A&D TM-2656 blood pressure monitor according to the British Hypertension Society Protocol. Blood Pressure Monitoring 2013, 18:223-226 | |
Andon Health Co. Ltd. | KD-391 | Clinic | Unknown Price | International Protocol | Independent | Wan y, Pan F, Liu Y, Liang Y, Yang Y, Zhao H & Xu Y. Validation of the Andon KD-391 semi-automated blood pressure monitor in adults according to the International Protocol. Blood Pressure Monitoring 2009, 14:181-184 | ||
Andon Health Co. Ltd. | KD - 551 (Derivative of KD5917) | Upper Arm | Unknown Price | international protocol 2010 | Independent | Wan-Gang Guo, Bing-Ling Li, Yong He, Yu-Sheng Xue, Hai-Yan Wang, Qiang-Sun Zheng and Ding-Cheng Xiang Validation of the Andon KD-5917 automatic upper arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010 Blood Pressure Monitoring, 2014, 19:242–245 |
||
Andon Health Co. Ltd. | KD - 5810 (Derivative of KD- 5917) | Upper Arm | Unknown Price | international protocol 2010 | Independent | Wan-Gang Guo, Bing-Ling Li, Yong He, Yu-Sheng Xue, Hai-Yan Wang, Qiang-Sun Zheng and Ding-Cheng Xiang Validation of the Andon KD-5917 automatic upper arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010 Blood Pressure Monitoring, 2014, 19:242–245 |
||
Andon Health Co. Ltd. | KD - 5811 (Derivative of KD- 5917) | Upper Arm | Unknown Price | international protocol 2010 | Independent | Wan-Gang Guo, Bing-Ling Li, Yong He, Yu-Sheng Xue, Hai-Yan Wang, Qiang-Sun Zheng and Ding-Cheng Xiang Validation of the Andon KD-5917 automatic upper arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010 Blood Pressure Monitoring, 2014, 19:242–245 |
||
Artsana | CS 410 | Clinic | Unknown Price | International Protocol | Independent | Pini C, Natalizi A, Gerosa PF, Frigerio M, Omboni S & Parati G. Validation of the Artsana CS 410 automated blood pressure monitor in adults according to the International Protocol of the European Society of Hypertension. Blood Pressure Monitoring 2008, 13:177-182 | ||
Artsana | CS 610 | Clinic | Unknown Price | international Protocol | Independent | Pini C, Pastori M, Bacceschi J, Omboni S & Parati G. Validation of the Artsana CS610 automated blood pressure monitor in adults according to the International Protocol of the European Society of Hypertension. Blood Pressure Monitoring 2007, 12:179-184 | ||
BTL | BTL-08 ABPM II (OEM derivative of Meditech ABPM-05) | ABPM | Unknown Price | BHS B/B | Independent | Barna I, Keazei A & Dunai A. Evaluation of Meditech ABPM-04 ambulatory blood pressure measuring device according to the British Hypertension Society protocol. Blood Pressure Monitoring 1998, 3:363-368 | ||
Dinamap | Procare 100 | Clinic | Over £100 | International Protocol | Independent | 6VDC rechargeable battery | Reinders A, Reggiori F & Shennan A. Validation of the Dinamap ProCare blood pressure device according to the International Protocol in an adult population. Blood Pressure Monitoring 2006, (11) 5: 293-296 | |
Dinamap | Procare 200 (Derivative of Dinamap 100) | Clinic | Over £100 | International Protocol | Independent | Reads BP & temp. 6VDC rechargeable batter | Reinders A, Reggiori F & Shennan A. Validation of the Dinamap ProCare blood pressure device according to the International Protocol in an adult population. Blood Pressure Monitoring 2006, (11) 5: 293-296 | |
Dinamap | Procare 230 (Derivative of Procare 100) | Clinic | Over £100 | International Protocol | Independent | Reinders A, Reggiori F & Shennan A. Validation of the Dinamap ProCare blood pressure device according to the International Protocol in an adult population. Blood Pressure Monitoring 2006, (11) 5: 293-296 | ||
Dinamap | Procare 420 (Derivative of Procare 100) | Clinic | Over £100 | International Protocol | Independent | Reinders A, Reggiori F & Shennan A. Validation of the Dinamap ProCare blood pressure device according to the International Protocol in an adult population. Blood Pressure Monitoring 2006, (11) 5: 293-296 | ||
EDAN | SA - 10 | Clinic | Unknown Price | international Protocol | Independent | Please note that this machine only passed the static part of ISO, not for ABPM | Validation of the EDAN SA-10 ambulatory blood pressure monitor in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard Bing Wang, Kui Liu, Linyi Li, Shimin Yin, and Jie Ren Blood Pressure Monitoring: August 2022 - Volume 27 - Issue 4 - p 276-279 doi: 10.1097/MBP.0000000000000600 |
|
EnvitecC-Wismar | PhysioQuant | ABPM | Unknown Price | International Protocol | Independent | Minor protocol violation; only 2 observers used in the validation study | Langewitz W & Tanner S. Validation of the PhysioQuant blood pressure measuring monitor in adults. Blood Pressure Monitoring 2009, 14(5): 220-221 | |
Ergoline | Ergoscan (Derivative of EnvitecWismar PhysioQuant) | ABPM | Unknown Price | International Protocol | Independent | Minor protocol violation; only 2 observers used in the validation study | Langewitz W & Tanner S. Validation of the PhysioQuant blood pressure measuring monitor in adults. Blood Pressure Monitoring 2009, 14(5): 220-221 | |
ERKA Kallmeyer Medizintechnik | Erkameter 125 PRO | Clinic | Unknown Price | international protocol 2010 | Independent | B Beime, C Deutsch, R Kruger, E Zimmermann and P Bramlage. Validation of the blood pressure measurement device Erkameter 125 PRO according to the European Society of Hypertension International Protocol revision 2010. Blood Pressure Monitoring 2016, 21:356-360 | ||
ET Medical Devices | Cardioline Walk200B (OEM derivative of IEM Mobil-O-Graph) | ABPM | Unknown Price | BHS A/A, International Protocol | Independent | Franssen PML & Imholz BPM. Evaluation of the Mobil-O-Graph new generation ABPM device using the ESH criteria. Blood Pressure Monitoring 2010, 15: 229-231. Weiss W, Tolle M, Zidek W & van der Giet M. Validation of the Mobil-O-Graph: 24h-blood pressure measurement device. Blood Pressure Monitoring, 15:225-228 | ||
GE | Carescape V100 (Derivative of Procare 100) | Clinic | Over £100 | International Protocol | Independent | 6VDC rechargeable battery | Reinders A, Reggiori F & Shennan A. Validation of the Dinamap ProCare blood pressure device according to the International Protocol in an adult population. Blood Pressure Monitoring 2006, (11) 5: 293-296 | |
GE Healthcare | Tonoport V | ABPM | Unknown Price | International Protocol | Independent | 2 x AA batteries | Haensel A, Utech K & Langewitz W. Validation of Tonoport V blood pressure measuring monitor in adults. Journal of Human Hypertension 2005, 0:1-6 | |
Healthworks | SCLV-2007 Cardio-Vascular Lab (Derivative of Spengler KP-7500D) | Clinic | Over £100 | International Protocol | Independent | USB driven | ABI, stroke volume & cardiac output | Barna I, Keazei A & Dunai A. Evaluation of Meditech ABPM-04 ambulatory blood pressure measuring device according to the British Hypertension Society protocol. Blood Pressure Monitoring 1998, 3:363-368 |
HeathSTATS International | Bpro | ABPM | Over £100 | International Protocol | Independent | 3C CR 2032 3V coin battery | Wrist device | Nair D, Tan S-Y, Gan H-W, Lim S-F, Tan J, Zhu M, Gao H, Chua N-H, Peh W-I & Mak K-H. Research letter: The use of ambulatory tonometric radial arterial wave capture to measure ambulatory blood pressure: the validation of a novel wrist-bound device in adults. Journal of human Hypertension 2008, 22:220-222 |
Heine Gamma | G7 (G5) | Aneroid | Unknown Price | International Protocol | Independent | Dorigatti F, Bonso E, Zanier A & Palatini P. Validation of Heine Gamma G7 (G5) and XXL-LFR aneroid devices for blood pressure measurement. Blood Pressure Monitoring 2007, 12:29-33 | ||
Heine Gamma | XXL-LF | Aneroid | Unknown Price | International Protocol | Independent | Dorigatti F, Bonso E, Zanier A & Palatini P. Validation of Heine Gamma G7 (G5) and XXL-LFR aneroid devices for blood pressure measurement. Blood Pressure Monitoring 2007, 12:29-33 | ||
Hingmed | WBP-02A | ABPM | Unknown Price | international protocol 2010 | Independent | Validation of the Hingmed WBP-02A device for ambulatoryblood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Claudio Fania, Leonardo Vezzù, Ilaria Lazzaretto and Paolo Palatini. Blood Pressure Monitoring 2019, 24:151–154 |
||
HoMedics | WGNBPA-540 | Clinic | Unknown Price | BHS A/A | Independent | Yarrows S. Accuracy of the HoMedics Walgreens Deluxe Arm Blood pressure device (WGNBPA-540) compared with auscultatiion in 85 individuals. Blood Pressure Monitoring 2001, 16:200-202 | ||
IEM | Mobil-O-Graph NG 24/48h ABPM | ABPM | Over £100 | BHS A/A, International Protocol | Independent | Software included. 2 x AA rechargeable charger included | Franssen PML & Imholz BPM. Evaluation of the Mobil-O-Graph new generation ABPM device using the ESH criteria. Blood Pressure Monitoring 2010, 15: 229-231. Weiss W, Tolle M, Zidek W & van der Giet M. Validation of the Mobil-O-Graph: 24h-blood pressure measurement device. Blood Pressure Monitoring, 15:225-228 | |
InBody | BPBIO250 | Clinic | Unknown Price | International Protocol | Independent | Kollias A, Gravvani A, Anagnostopoulos I, Kyriakoulis K, Bountzona I, Menti A and Stergiou G. Validation of the InBody BPBIO250 oscillometric blood pressure monitor for professional office use in general population according to the Associatio for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard. Blood Pressure Monitoring 2020, 25:115–117 |
||
InBody | BPBIO320 | Waiting Room | Unknown Price | International Protocol | Independent | Kiosk device for public use | Kollias A, Stamolliu E, Kyriakoulis KG, Papadatos SS, Stergiou, GS. Validation of the single-cuff oscillometric blood pressure monitor InBody BPBIO320 for public use according to the 2010 European Society of Hypertension International Protocol. Blood Pressure Monitoring 2019, 24:30-32 |
|
InBody | BPBIO750 (Derivative of BPBIO320) | Waiting Room | Unknown Price | international protocol 2010 | Independent | Kiosk device for public use | Kollias A, Stamolliu E, Kyriakoulis KG, Papadatos SS, Stergiou, GS. Validation of the single-cuff oscillometric blood pressure monitor InBody BPBIO320 for public use according to the 2010 European Society of Hypertension International Protocol. Blood Pressure Monitoring 2019, 24:30-32 |
|
Keeler | Big Ben Square, Round , Windsor, Palm, Classic (All derivative of Riester Big Ben Square) | Aneroid | Unknown Price | International Protocol | BHS | Tested and approved by the BIHS Validation Service | ||
Medaval | QMon-20 | Clinic | Unknown Price | ANSI/ESH/ISO 81060–2:2018 | Independent | Sergio Rico-Martín, Marisol Sánchez-Bacaicoa, Julián F. Calderón-García, Pedro J. Labrador-Gómez, Jorge M. De Nicolás Jiménez, Juan Villa-Rincón, Nicolás R. Robles, Carlos Guijarro, Chiara Iammarino, Enrique Rodilla-Salas and Juan F. Sánchez Muñoz-Torrero Validation of the QMon-20 oscillometric blood pressure monitor for professional office use in the general population according to the ANSI/ESH/ISO 81060–2:2018 protocol Blood Pressure Monitoring 2021, 26:393–395 |
||
Meditech | ABPM-04 | ABPM | Over £100 | BHS B/B | Independent | 4 x AA Alkaline or rechargeable | Barna I, Keazei A & Dunai A. Evaluation of Meditech ABPM-04 ambulatory blood pressure measuring device according to the British Hypertension Society protocol. Blood Pressure Monitoring 1998, 3:363-368 | |
Meditech | ABPM-05 (Derivative of ABPM-04) | ABPM | Over £100 | BHS B/B | Independent | Barna I, Keazei A & Dunai A. Evaluation of Meditech ABPM-04 ambulatory blood pressure measuring device according to the British Hypertension Society protocol. Blood Pressure Monitoring 1998, 3:363-368 | ||
Meditech | ABPM-06 | ABPM | Over £100 | BHS A/A | Independent | Validated and approved against the BHS protocol only. Protocol violations noted for ESH-IP2 and IOS | Hermányi Z, Pokoly B, Visolyi G, Barna I. Evaluation of Meditech ABPM-06 ambulatory blood pressure measuring device, according to the European Society of Hypertension, the British Hypertension Society and the International Organization for Standardization protocol. Blood Press Monit. 2019; 24:208–211 |
|
Medset Medizintechnik | Scanlight III (OEM derivative of IEM Mobil-O-Graph) | ABPM | Unknown Price | BHS A/A, International Protocol | Independent | Franssen PML & Imholz BPM. Evaluation of the Mobil-O-Graph new generation ABPM device using the ESH criteria. Blood Pressure Monitoring 2010, 15: 229-231. Weiss W, Tolle M, Zidek W & van der Giet M. Validation of the Mobil-O-Graph: 24h-blood pressure measurement device. Blood Pressure Monitoring, 15:225-228 | ||
Microlife | WatchBP Home | Pregnancy | Under £100 | BHS A/A | Independent | L-XL cuff available. 4 x AA batteries mains adaptor available | AF detection in this device does not mean that this device is validated for use in patients with AF | Chung Y, de Greeff A & Shennan A. Valdiation and compliance of a home monitoring device in pregnancy: Microlife Watch BP Home. Hypertension in Pregnancy, 28, 3:348-359. |
Microlife | WatchBP 03 | ABPM | Over £100 | International Protocol | Independent | PC link. 4 x AAA Alkaline Batteries | AF detection in this device does not mean that this device is validated for use in patients with AF | Ragazzo F, Saladini F & Palatini P. Validation of the Microlife WatchBP 03 device for clinic, home and ambulatory blood pressure measurement, according to the International Protocol. Blood Pressure Monitoring 2010, 15(1):59-62 |
Microlife | A3PC | Clinic | Unknown Price | ANSI/AAMI/ISO 81060-2:2013 | Independent | approved for use in diabetes, but with medium large cuff only | Validation of the Microlife BP A3 PC upper arm blood pressure monitor in patients with diabetes mellitus according to the ANSI/AAMI/ISO 81060-2:2013 protocol Beate Beimea; Ralf Krügera; Gertrud Hammelb; Peter Bramlagea; Cornelia Deutscha Blood Press Monitoring: February 2018; 23(1):52-57. doi: 10.1097/MBP.0000000000000302 |
|
Microlife | B3AFIB | Upper Arm | Unknown Price | ANSI/AAMI/ISO 81060-2:2013 | Independent | Beime, Beate; Bramlage, Carsten; Krüger, Ralf; Deutsch, Cornelia; van Mark, Gesine; Bramlage, Peter; Botta, Beate Validation of the Microlife BP B3 AFIB upper arm blood pressure monitor in adults and adolescents according to the ANSI/AAMI/ISO 81060-2:2019 protocol Blood Pressure Monitoring: August 2021 - Volume 26 - Issue 4 - p 299-304 doi: 10.1097/MBP.0000000000000530 |
||
Microlife | WatchBP Office | Clinic | Unknown Price | International Protocol | Independent | Bluetooth, 3 function modes allowing simultaneous measurement in both arms. | Stergiou GS, Tzamournanis D, Protogerou A, Nasothimiou E & Kapralos C. Validation of the Microlife WatchBP Office professional device for office blood pressure measurement according to the International Protocol. Blood Pressure Monitoring 2008, 13:299-303. | |
Microlife | WatchBP Office ABI | Clinic | Unknown Price | International Protocol | Independent | Validated for upper arm use only. NB AF detection does not mean that this device has been validated for use in patients with AF | Saladini F, Benetti E, Masiero S & Palatini P. Accuracy of Microlife WatchBP Office ABI monitor assessed according to the 200 European Society of Hypertension protocol and the British Hypertension Society protocol. Blood Pressure Monitoring 2011, 16:258-261 |
|
Microlife | WatchBP Office ABI | Obese Adults | Unknown Price | BHS A/A | Independent | Validated for upper arm use only. NB AF detection does not mean that this device has been validated for use in patients with AF | Masiero S, Saladini F, Benetti E & Palatini P. Accuracy of the Microlife large-extra large sized cuff (32-52cm) coupled to an automatic oscillometric device. Blood Pressure Monitoring 2011, 16:99-102 | |
Microlife | 3AS1-2 | Pregnancy | Unknown Price | BHS B/A, BHS A/A, International Protocol | Independent | Semi-automatic. 2 x AAA batteries | Validated for use in pregnancy only. An insufficient population of pre-eclamptic patients was used in the validation study to inform recommendation. Minor protocol violation; only 2 observers used in the validation study. Validated in pregnant population with low blood pressure in a low resource settings in low and middle income countries | de Greef A, Nathan , Stafford N, Liu B & Shennan A. Development of an acccurate oscillometric blood pressure device for low resource settings. Blood Pressure Monitoring 2008, 13:342-348. Nathan HL, de Greeff A, Hezelgrave NL, Chappell LC and Shennan AH. An accurate semiautomated oscillometric blood pressure device for use in pregnancy (including pre-eclampsia) in a low-income and middle-income country populaton: the Microlife 3AS1-2. Blood Pressure Monitoring 2015, 20:52-55. Nathan, L, de Greeff A, Hezelgrave NL, Chappell LC and Shennan A. Accuracy of the Microlife 3AS1-2 blood pressure device in a pregnant population with low blood pressure. Blood Pressure Monitoring 1025, 20:299-302 |
Mindray | Accutor V (Derivative of Mindray Accutor Plus) | Clinic | Over £100 | BHS A/A | Independent | Child, Standard Adult & Large Adult Cuffs: 12v rechargeable SLA or 10.8v rechargeable lithiam iron battery | Anwar YA, Tendler BE, McCabe EJ et al. Evaluation of the Datascope Accutorr Plus according to the Association for Advancement of Mdical Instrumentation. Blood Pressure Monitoring. 1997, 2:105-110 | |
Mindray | Accutor V (Derivative of Mindray Accutor Plus) | Children | Over £100 | International Protocol | Independent | Child, Standard Adult & Large Adult Cuffs: 12v rechargeable SLA or 10.8v rechargeable lithiam iron battery | Tested in normotensive children only. No evidence of accuracy in hypertensive children. | Wong SN, Sung RYT & Leung LCK. Validation of three oscillometric blood pressure devices against auscultatory Mercury sphygmomanometers in children. Blood Pressure Monitoring 2006, 11:281-291 |
Mindray | Accutor Plus | Clinic | Over £100 | BHS A/A | Independent | Child, Standard Adult & Large Adult Cuffs: 12v rechargeable SLA or 10.8v rechargeable lithiam iron battery | Anwar YA, Tendler BE, McCabe EJ et al. Evaluation of the Datascope Accutorr Plus according to the Association for Advancement of Mdical Instrumentation. Blood Pressure Monitoring. 1997, 2:105-110 | |
Mindray | Accutor Plus | Children | Over £100 | International Protocol | Independent | Child, Standard Adult & Large Adult Cuffs: 12v rechargeable SLA or 10.8v rechargeable lithiam iron battery | Tested in normotensive children only. No evidence of accuracy in hypertensive children. | Wong SN, Sung RYT & Leung LCK. Validation of three oscillometric blood pressure devices against auscultatory Mercury sphygmomanometers in children. Blood Pressure Monitoring 2006, 11:281-291 |
Mindray | BeneView T5 | Vital Signs Monitor | Unknown Price | International Protocol | BIHS | Validated for BP measurement only | Tested and approved by the BHS Validation Service | |
Mindray | BeneVision N12 (Derivative of BeneView T5) | Vital Signs Monitor | Uknown Price | International Protocol | BIHS | Validated for BP measurement only | Tested and approved by the BHS Validation Service | |
Mindray | iMEC12 (Derivative of Mindray BeneView T5) | Vital Signs Monitor | Unknown Price | International Protocol | BIHS | Validated for BP measurement only | Tested and approved by the BHS Validation Service | |
Mindray | iPM 12 (Derivative of Mindray BeneView T5) | vital Signs Monitor | Unknown Price | International Protocol | BIHS | Validated for BP measurement only | Tested and approved by the BHS Validation Service | |
Mortara (formerly Tibia Medical) | Ambulo 2400 | ABPM | Over £100 | BHS A/A | Independent | 2 x AA Alkaline or NiMH batteries | Alpert BS. Validation of the Tiba Medical Ambulo 2400 ambulatory blood pressure monitor to the ISO and British protocol. Blood Pressure Monitoring 2010, 15:275-277 | |
Nissei | DM-3000 | Clinic | Over £100 | International Protocol | Independent | Auscultatory & Oscillometric function. Rechargeable battery pack. AC mains adaptor included | Validated for use in automatic oscillometric mode and in blinded auscultatory mode at default deflation rate of 2.5mmHg/sec | Tasker F, de Greef A & Shennan A. Development and validation of a blinded hyprid device according to the European Hypertension Society Protocol: Nissei DM-3000. Journal of Human Hypertension (2010), doi: 10.1038/jhh.2009.113 |
Novacor | Diasys 3 | ABPM | Over £100 | AAMI/ANSI/ISO 81060-2:2013 | Independent | Atkins N, Fania C and Palatini P. Validation of the blood pressure measurement technolog used in the Novacor Diasys 3 (DIS 0001-00) upper arm device for ambulatory blood pressure measurement, according to the requirements of the AAMI/ANSI/ISO 81060-2:2013 standard (for both a general study and a cardiac-stress study in adults) and of the European Society of Hypertension International Protocol revision 2010. Blood Pressure Monitoring 2021, 26:70–77 |
||
Novacor | Diasys 3 Plus | ABPM | Over £100 | international protocol 2010 | Independent | ABPM device | Approved according to the International Protocol (2010) | Atkins N, Fania C and Palatini P. Validation of the blood pressure measurement technology used in the Novacor Diasys 3 Plu (DIP-0001-00) upperarm device for ambulatory blood pressure measurement, according to AAMI/ANSI/ISO 81060-2:2013, ESH-IP 2010 and MEDDEV 2.7/1. Blood Pressure Monitoring 2020, 25:359–367 |
Omron | M7 (HEM-780-E) | Obese Adults | Over £100 | International Protocol | Independent | One-size cuff 22-42cm, 4 x AA batteries, AC mains adapter available | Also validated for use in pregnancy - See separate note | El Feghali RN, Topouchian JA, Pannier BM, El Assaad HA & Asmar RG. Validation of the Omron M7 (HEM0780-E) blood pressure measuring devbice in a population requiring large cuff use according to the International Protocol of the European Society of Hypertension. Blood Pressure Monitoring 2007, 12: 173-178. |
Omron | M7 (HEM-780-E) | Pregnancy & Pre-eclampsia | Over £100 | International Protocol | Independent | One-size cuff 22-42cm, 4 x AA batteries, AC mains adapter available | Also validated for use in obese adults - See separate note | de Greef A, Beg Z, Gangji Z, Dorney E & Shennan A. Accuracy of inflationary versus deflationary oscillometry in pregnancy and pre-eclampsia; Omron MIT versus Omron M7 Blood Pressure Monitoring 2009, 14:37-40 |
PAR Medizintechnik | Physio Port (Derivative of Tonoport VI) | ABPM | Unknown Price | International Protocol 2010 | Independent | Abou-Dakn M, Dohmen C and Wenzel S. Validation of the Tonoport VI ambulatory blood pressure monitor in adults according to the European Society of Hypertension International Protocol revision 2010. Journal of Human Hypertension (2016), 1-4 | ||
PAR Medizintechnik | Tonoport VI | ABPM | Unknown Price | International Protocol 2010 | Independent | Abou-Dakn M, Dohmen C and Wenzel S. Validation of the Tonoport VI ambulatory blood pressure monitor in adults according to the European Society of Hypertension International Protocol revision 2010. Journal of Human Hypertension (2016), 1-4 | ||
Reister | RBP-100 (Derivative of Microlife WatchBP Office ABI) | Clinic | Unknown Price | International Protocol | Independent | Validated for upper arm use only. NB AF detection does not mean that this device has been validated for use in patients wit AF | Saladini F, Benetti E, Masiero S & Palatini P. Accuracy of Microlife WatchBP Office ABI monitor assessed according to the 200 European Society of Hypertension protocol and the British Hypertension Society protocol. Blood Pressure Monitoring 2011, 16:258-261 |
|
Riester | Big Ben Square | Aneroid | Unknown Price | International Protocol | BIHS | Tested and approved by the BHS Validation Service | ||
Riester | Big Ben Round (Derivative of Big Ben Square) | Aneroid | Unknown Price | International Protocol | BIHS | Tested and approved by the BHS Validation Service | ||
Riester | E-mega (Derivative of Big Ben Square) | Aneroid | Unknown Price | International Protocol | BIHS | Tested and approved by the BHS Validation Service | ||
Riester | exacta (Derivative of Big Ben Square) | Aneroid | Unknown Price | International Protocol | BIHS | Tested and approved by the BHS Validation Service | ||
Riester | Minimus II (Derivative of Big Ben Square) | Aneroid | Unknown Price | International Protocol | BIHS | Tested and approved by the BHS Validation Service | ||
Riester | Minimus III (Derivative of Big Ben Square) | Aneroid | Unknown Price | International Protocol | BIHS | Tested and approved by the BHS Validation Service | ||
Riester | Precisa N Shock-proof (Derivative of Big Ben Square) | Aneroid | Unknown Price | International Protocol | BIHS | Tested and approved by the BHS Validation Service | ||
Riester | precisa N (Derivative of Big Ben Square) | Aneroid | Unknown Price | International Protocol | BIHS | Tested and approved by the BHS Validation Service | ||
Riester | R1 shock-proof (Derivative of Big Ben Square) | Aneroid | Unknown Price | International Protocol | BHS | Tested and approved by the BHS Validation Service | ||
Riester | ri-cardio (Derivative of SunTech Medical Oscar 2) | ABPM | Unknown Price | BHS A/A, International Protocol | Independent | Goodwin J, Bilous M, Winship S, Finn P & Jones SC. Validation of the Oscar 2 oscillometric 24-h ambulatory blood pressure monitor according to the British Hypertension Society protocol. Blood pressure Monitoring 2007, 12:113-117. Jones SC, Bilous M, Winship S, Finn P & Goodwin J. Validation of the Oscar 2 oscillometric 24-hour ambulatory blood pressur monitor according to the International Protocol for the validation of blood pressure measuring devices. Blood Pressure Monitoring 2004,9(4) :219-223 |
||
Riester | ri-medic (Derivative of SunTech 247) | Vital Signs Monitor | Unknown Price | BHS A/A, International Protocol | Independent | Friz HP, Facchetti R Primitz L, Beltrame L, Glbiati V, Bicioppo A, Bombelli M & Sega R. Simultaneous validation of the SunTech 247 diagnostic station blood pressure measurement device according to the British Hypetension Society protocol, the International Protocol and the Association of Medical Instrumentation standards. Blood Pressure Monitoring 2009, 14:222-227 | ||
Riester | ri-san (Derivative of Big Ben Square) | Aneroid | Unknown Price | International Protocol | BHS | Tested and approved by the BHS Validation Service | ||
Schiller | BR-102 Plus | ABPM | Over £100 | International Protocol | Independent | 2 x rechageable AA batteries | Denchev SV, Simova II & Matveev MG. Evaluation of the Schiller BR-102 Plus non-invasive ambulatory blood pressure monitor according to the International Protocol introduced by the Working Group on Blood Pressure Monitoring of the European Society of Hypertension. Blood Pressure Monitoring 2007, 12:329-333 | |
SOMNOmedics | SOMNOmedics ABPMpro | Upper Arm | Unknown Price | International Protocol | Independent | Validated up to an upper arm circumference of 39 cm | Roth B, Bothe TL, Patzak A, Pilz N. Validation of the ABPMpro ambulatory blood pressure monitor in the general population according to AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018). Blood Press Monit. 2023 Jun 1;28(3):158-162. doi: 10.1097/MBP.0000000000000640. Epub 2023 Apr 5. PMID: 37016982; PMCID: PMC10132455. |
|
Sorin | Agilis | ABPM | Over £100 | International Protocol | Independent | Software included. 2 x AA Alkaline or NiMH rechargeable | Mallion JJM, Pierre H, Neuder Y, Ormezzano O & Baguet JP. Validation of the Agilis ambulatory blood pressure monitor according to the European Society of Hypertension Intewrnational Protocol for validation of blood pressure measuring devices in adults. Blood Pressure Monitoring 2005, 10(2):97-101 | |
Spacelabs | SL90207 | ABPM | Over £100 | BHS B/B | Independent | Software and download cable included. 4 x AA batteries | O’Brien E, Mee F, Atkins N & O’Malley K. Accuracy of the SpaceLabs 90207 determined by the British Hypertension Society Protocol. J Hypertens 1991, 9 (suppl 5): S25-S31 | |
Spacelabs | SL 90217 | ABPM | Over £100 | BHS A/A | Independent | Sofware & download cable included. Ultralight. 3 x AA batteries | Baumgart P & Kamp J. Accuracy of the SpaceLabs Medical 90217 ambulatory blood pressure monitor. Blood Pressure Monitoring 1998, 3:303-307 | |
Spacelabs | 90227 OnTrak | ABPM | Unknown Price | BHS A/A | Independent | Approved in adults only via BHS Protocol. Not Approved: ESH IP2 Protocol violation re BP distribution. AAMI Protocol violation re subject inclusion. | de Greef A and Shennan A. Validation of the Spacelabs 90227 OnTrak device according to the European and British Hypertension Societies as well as the American protocols. Blood Pressure Monitoring. 2020, 25:110–114 |
|
Spengler | Pro M | Upper Arm | Unknown Price | International Protocol | Independent | de Greef A, Ahmed N, Sheppard E & Shennan A. A self-measurement device suitable for hospital use: validation of the Spengler Pro M according to the International Protocol of the European society of Hypertension. Blood Pressure Monitoring 2008, 13:183-186 |
||
Suntech | Oscar 2 | ABPM | Uknown Price | BHS A/A, International Protocol | Independent | 2 x AA Alkaline or NiMH rechargeable | Goodwin J, Bilous M, Winship S, Fin P & Jones SC, Calidtion of the Oscar 2 oscillometric 24-h ambulatory blood pressure monitor according to the British Hypertension Society Protocol. Blood Pressure Monitoring, 2007, 12:113-117. Jones SC, Bilou M, WinshipS, Finn P & Goodwin J. Validation of the OSCAR 2 oscillometric 24-hour ambulatory blood pressure monitor according to the International Protocol for the validation of blood pressure monitoring devices in adults. Blood Pressure Monitoring 2004, 9(4):219-225. | |
Unimedic | Tensioday | ABPM | Over £100 | BHS A/A | Independent | 4 x AA batteries | ||
Uscom | BP+ (Derivative of Tonoport VI) | Clinic | Over £100 | international protocol 2010 | Independent | AC power supply & 3 cuffs included (Small Adult, Adult & Large Adult Cuffs), Graphic display with waveforms | In the absence of a recognised protocol for testing central pressure accuracy, this device is validated for brachial measurement only | Abou-Dakn M, Dohmen C and Wenzel S. Validation of the Tonoport VI ambulatory blood pressure monitor in adults according to the European Society of Hypertension International Protocol revision 2010. Journal of Human Hypertension (2016), 1-4 |
Welch Allyn | ABPM 6100 (Derivative of Suntech Oscar 2) | ABPM | Over £100 | BHS A/A | Independent | 2 x AA batteries | Goodwin J, Bilous M, Winship S, Fin P & Jones SC. Validation of the Oscar 2 oscillometric 24-h ambulatory blood pressure monitor according to the British Hypertension Society protocol. Blood Pressure Monitoring 2007, 12:113-117 | |
Welch Allyn | ABPM-7100 (Derivative of Mobil-O-Graph NG 24/48H ABPM) | ABPM | Unknown Price | BHS A/A | Independent | Validation of the Mobil-O-Graph 24 h-blood pressure measurement device. Weiss W, Tolle M, Zidek W & van der Giet M. Blood Pressure Monitoring 2010, 15:225-228 | ||
Welch Allyn | ProBP 2400 (Derivative of Microlife WatchBP Office ABI) | Clinic | Unknown Price | BHS A/A, International Protocol | Independent | Validated for upper arm use only | Saladini F, Benetti E, Masiero S and Palatini P. Accuracy of Microlife WatchBP Office ABI monitor assessed according to the 2002 European Society of Hypertension protocol and the British Hypertension Society protocol. Blood Pressure Monitoring 2011, 16:258-261 | |
Welch Allyn | ProBP 2400 (Derivative of Microlife WatchBP Office ABI) | Obese Adults | Unknown Price | BHS A/A | Independent | Validated for upper arm use only | Masiero S, Saladini F, Benetti E, and Palatini P. Accuracy of the Microlife large-extra large-sized cuff (32-52cm) coupled to an automatic oscillometric device. Blood Pressure Monitoring 2011, 16:99-102 | |
Welch Allyn | ProBP 3400 | Upper Arm | Unknown Price | BHS A/A | Independent | Mains Powered | Alpert B. Validation of the Welch Allyn ProBP 3400: a device for modern medical practice. Blood Pressure Monitoring 2011, 16:156-158 |
|
Withings | BPM Connect (WPM05) | Pregnancy & Pre-eclampsia | Over £100 | International Protocol | Independent | Zelveian P, Topouchian J, Hakobyan Z, Asmar J, Gharibyan H, Asmar R Clinical Accuracy of the Withings BPM Connect for Self-Blood Pressure Measurements in Pregnancy and Pre-Eclampsia: Validation According to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard Vascular Health and Risk Management, 2022, 18, 181–189 |
||
Withings | BPM Connect Pro (Derivative of BPM Connect WPM05) | Clinic | Unknown Price | International Protocol | Independent | Zelveian P, Topouchian J, Hakobyan Z, Asmar J, Gharibyan H, Asmar R Clinical Accuracy of the Withings BPM Connect for Self-Blood Pressure Measurements in Pregnancy and Pre-Eclampsia: Validation According to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard Vascular Health and Risk Management, 2022, 18, 181–189 |
||
YuWell | YE900 | Clinic | Unknown Price | International Protocol | Independent | Also validated for paediatric use in a clinical setting | Hui-Jie Zhang, Juan Zhang, Su-Lan Wang, Jin Zhang, Li-Ni Teng, Shu-Juan Zhang, De-Jun Zhou and Ming-Zhi Long. Validation of the YuWell YE900 oscillometric blood pressure monitor for professional office use in adults and children according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) Blood Pressure Monitoring 2021, 26:396–399 |